Thank you for writing to us. One of our executive will reach back to you through your submitted medium. In case there’s an urgency, feel free to connect over WhatsApp for faster response.
Prefer calling? Dial +9108048035791 (international callers) or 08048035791 (Indian callers).
SANSEL Model TCAL 1405 / -30 Dry well Temperature calibrator are high performance for industrial and Laboratory atmosphere and covering temperature range from -30* to 100°C. Which are ideally suited onsite calibration of thermocouple, RTDs, Indicator with sensors & 4-20mA signal transmitter with sensor & indicator in Food, Pharmaceutical and all types of process industries and Calibration Laboratories. Having it’s advantage of user friendly, light weight, portable, fast settling, better uniformity and accuracy over a wide temperature range. TCAL 1405 / -30 having interchangeable inserts to suit for calibration of different diameter of sensors. Additional Inserts (optional) with special drillings are available on request.
Technical Specifications :
Model: TCAL 1405 / - 30
Temperature Range: -30* to 100°C (* @ Ambient 25 °C)
Resolution: 0.01°C
Display: 3.5“ True TFT Multi colour touch screen display
Temperature readout: °C / °F
Block Stability: ± 0.03°C
Axial Uniformity upto 40mm: ± 0.06 °C (from -30 to 50 °C), ± 0.1 °C (from 51 to 100 °C)
Radial Uniformity: ± 0.05°C
Loading Effect: ± 0.04°C
Hysteresis: ± 0.02°C
Well Dimension (dia. x height x hole depth): 25 x 145 x 130 mm
Standard well holes in mm: 6.2 x 3nos, 8.1 x 1nos
Power supply: 230V AC ± 10% / 50Hz (Optional 110V AC), 500 W
Overall Dimension in mm: 142.0 x 233.5 x 271.5 (L x B x H)(Approx.)
Instrument Weight: 8 kg (Approx.)
The Timer Calibrator TFC-1901 is a precision instrument. It is Primarily designed to generate the accurate time intervals for calibrating time-related devices, ensuring traceability to National/International standards. Built with a internal 10MHz OCXO/TCXO time base Oscillator or External input Frequency from standard Lab source and also with Non contact RPM calibrating source all in a single unit — with potential-free MOSFET trigger output connector terminal to support an external Multi-stopwatch fixture (Model : TFC-19F) — to eliminate manual START/STOP timing error and 10MHz OCXO/TCXO Reference frequency output is provided on rear side terminal of Instrument.
Technical Specifications:
Model: TFC 1901
Display: 4.3” Multi-colour TFT touchscreen display.
Range: Digital Electronic Stopwatch: 100 ms to 86400.00000 s; Mechanical Stopwatch & Electronic Industrial Timer: 1 s to 86400.00000 s.
Resolution: 10 µs to 1 s (selectable: 0.00001 / 0.0001 / 0.001 / 0.01 / 0.1 / 1 s).
Display Format: Time in HH:MM:SS.fraction.
Mode 1 – Precision Stopwatch / Timer: High-resolution elapsed-time counter driven directly from the Precision internal oscillator reference, supporting free-run stopwatch and preset auto-stop timer operation with Master and DUC triggering.
Time Base: Directly referenced to the internal 10 MHz OCXO/TCXO.
Auto-Stop: Auto-stop at user-set target time.
Master vs. Calibration Mode: Master: touch control, outputs enabled, Auto-Stop available, Up/Down count; Calibration: external-input control, outputs disabled, input detection enabled, no Auto-Stop.
Accuracy: Digital Electronic Stopwatch – OCXO: ±(5.02 × 10⁻⁸ × Te + 2 ms); TCXO: ±(1.003 × 10⁻⁶ × Te + 2 ms); Mechanical Stopwatch & Electronic Industrial Timer – OCXO: ±(5.02 × 10⁻⁸ × Te + 5 ms); TCXO: ±(1.003 × 10⁻⁶ × Te + 5 ms), where Te = elapsed time.
Trigger I/O: Master generates START/STOP/RESET output pulses (30 µs) and continuous output; Calibration accepts external START/STOP/RESET, pulse mode and selectable continuous mode.
Mode 2 – Frequency Totalizer: Precision pulse totalizer that counts input events entirely in hardware, converts the count to elapsed time at the selected resolution, and simultaneously measures live input frequency referenced to the OCXO/TCXO.
Run Time: 24 hours continuous run at 100 kHz without rollover (1.728 × 10¹⁰ pulses).
Input / Counting: Dedicated pulse-count input with standard logic-level pulse train and fully hardware-based counting.
Frequency Measurement: Measured against OCXO/TCXO reference; range from 1 Hz to 100 kHz.
Resolution / Frequency Map: 0.00001 s ↔ 100 kHz, 0.0001 s ↔ 10 kHz, 0.001 s ↔ 1 kHz, 0.01 s ↔ 100 Hz, 0.1 s ↔ 10 Hz, 1 s ↔ 1 Hz.
Mode 3 – RPM Generator: Front-panel LED blinks at 50% duty cycle, matching the RPM value for visual RPM checking using a stroboscope or similar method.
Range for Non-contact: 0 to 204,000 RPM.
Output / Conversion: Front-panel LED output, 50% duty-cycle square wave; f (Hz) = RPM / 60.
Model : TCAL 1503 / -40
Display : 4.3” True colour TFT display with touch screen
Temperature Range : -40 to 120°C @23°C Ambient
Resolution : 0.01 °C
Block Stability : Better than ± 0.03°C at full range
Block Radial Uniformity : ± 0.02°C at full range
Block Axial Uniformity (upto 40mm) : ± 0.18°C at -45°C
± 0.1°C at 0°C
± 0.05°C at 50°C
± 0.11°C at 100°C
± 0.17°C at 120°C
Temperature Read out : °C / °F (selectable)
Loading effect: Better than ± 0.08°C
hysteresis: Better than ± 0.03°C
Cooling time :
30 mins (25 to -45 °C)
15 mins (25 to -30 °C)
65 mins (120 to 25 °C)
Heating time : 25 mins (25 to 120 °C)
75 mins (-45 to 120 °C)
52 mins (-45 to 25 °C)
Stabilization time: 10 minutes
Insert (Dia. x Height): 30 x 160 mm
Test sensor immersion depth: 130 mm
Standard well holes in mm: Insert No. 1 → 5.1, 6.2, 6.2 & 8.1, Insert No. 2 → 6.2, 6.2 & 10.2
Measure inputs: RTD-PT 100, PT 1000 / TC - J, K, R, S, T, E, B , N , Analog – mA, V, mV, ohms
Cold junction Accuracy: ± 0.3°C
UUC Sensor supply input: 24V DC ± 10% / 50mA max.
Power supply: 230V AC ± 10% / 50Hz (optional 110V AC)
Power Consumption: 500 VA
Switch Test : Provided
Overall Dimension in mm: 390 x 185 x 356 (L x B x H) (Approx.)
Weight: 17.3 kg (with well insert) (Approx.)
Model : TCAL 1503 / -100
Display : 4.3” True colour TFT display with touch screen
Temperature Range : -100°C to 40°C @20°C to 30°C Ambient
Resolution : 0.01 °C
Block Stability (30 mins) : Better than ± 0.03°C at full range
Block Radial Uniformity : ± 0.05 °C at full range
Temperature Read out : °C / °F (selectable)
Heating time :
25°C to 40°C : 30 min
-100 to 25°C : 80 min
Cooling Time: 25°C to -100°C : 85 min
40 to 25°C : 12 min
Stabilization time : 15 minutes
Insert Dia. x Height: 30 x 120 mm
Test sensor immersion depth: 115 mm
Standard well holes in mm : Insert No. 1 : 5.1, 6.2, 6.2 & 8.1
Insert No. 2 : 6.2, 6.2 & 10.2
Measure inputs: RTD – PT 100, PT 1000, TC – J, K, R, S, T, E, B, N , Analog – mA, V, mV, Ohms
Cold junction Accuracy: ± 0.3°C
Communication: RS 485 MODBUS
Switch Test: Provided
Sensor supply output: 24V DC ± 10% / 50mA max.
Power supply: 230V AC ± 10% / 50Hz (Optional 110V AC)
Power Consumption : 500 VA
Overall Dimension in mm : 356 x 210 x 534 (L x B x H)(Approx.)
Instrument Weight :15.9 kg(with well insert)(Approx.)